ORCID
Emad El-Giar: https://orcid.org/0000-0001-9928-7758; David Wipf: https://orcid.org/0000-0003-2365-1175
MSU Affiliation
College of Arts and Sciences; Department of Chemistry
Creation Date
11-1-2007
Creation Date
2026-11-13
Abstract
In this paper, we present a simple, relatively inexpensive, and reproducible method for the construction of iridium oxide (IrOx)-pH sensitive ultramicroelectrodes (UMEs). In this method, the pH-sensing layer is based on an IrOx film formed by potential cycling of iridium microparticles (diameter < 10 μm) in 0.5 M H2SO4 at 3.0 V/s. The IrOx-pH microsensors prepared this way are chemically stable, uniform, and exhibit rapid and linear response to H+ over the range 2-12 pH with a super-Nernstian slope (-64 ± 7 mV/pH) at room temperature. The pH response is reproducible and remains stable over a period of two months. Coexistence of redox ions such as Fe(CN)63- and Ru(NH3)63+ has little influence on the determination, however Ni2+ does produce an apparent pH shift. The application of these pH-sensitive UMEs is demonstrated in imaging the pH changes due to the oxygen reduction reaction at the surface of a cast iron sample.
Keywords
pH sensing, metal oxide electrodes, ion-selective pH UME, iridium oxide (IrOx), potential cycling, SECM, cast iron, oxygen reduction reaction
Publication Date
11-1-2007
Publication Title
Journal of Electroanalytical Chemistry
Publisher
Elsevier; International Society of Electrochemistry (ISE)
First Page
147
Last Page
154
Creative Commons License

This work is licensed under a Creative Commons Attribution-NonCommercial-No Derivative Works 4.0 International License.
Rights
© 2007 Elsevier B.V. All rights reserved.
Recommended Citation
El-Giar, E. E.-D. M., & Wipf, D. O. (2007). Microparticle-based iridium oxide ultramicroelectrodes for pH sensing and imaging. Journal of Electroanalytical Chemistry, 609(2), 147–154. https://doi.org/10.1016/j.jelechem.2007.06.022